Accuracy and precision
On this page
A clock that is always five minutes fast agrees with itself every day, and it is wrong every day. A clock that is sometimes fast and sometimes slow may be right on average, and you cannot trust any one look at it. Instruments have the same two faults. Precision is how closely repeated readings agree with each other, and accuracy is how close a reading is to the true value, which needs a small offset as well as good precision.
See it
A pressure transmitter is compared with a trusted reference gauge that shows 4.000 bar. Ten readings of the transmitter lie between 4.01 and 4.05 bar. You take 1,000 more readings and average them. What will the average be close to?
The centre of the target and the dashed line on the scale are the true value, 4.000 bar from a reference gauge. A shot to the right is a reading that is too high. The bias moves the whole group, and the spread makes the group wider. The figure opens on the transmitter of the worked number.
The idea
Measure one steady quantity ten times. The readings differ a little from each other, and their mean differs a little from the true value. These are two separate faults.
The bias is the amount by which the mean is wrong. It has a sign and a unit, and a known bias can be subtracted. An instrument with a small bias is called true.
the reference value comes from a better instrument, because the true value is never known exactly
The spread is the scatter of the readings around their own mean, expressed as a standard deviation. A small spread is a high precision. Precision says nothing about the true value.
A reading is accurate only when the instrument is both true and precise. A maker's figure such as ±0.05 bar is a limit that the error of one reading should not pass. It is not the error itself.
One worked number
A reference gauge shows 4.000 bar. Ten readings of a transmitter on the same pipe are 4.03, 4.02, 4.04, 4.03, 4.01, 4.03, 4.05, 4.02, 4.03 and 4.04 bar. What is the bias? How close are the readings to the reference after the bias is subtracted?
Show the worked solution, 3 steps
- Find the mean
The ten readings add to 40.30 bar, so the mean is 40.30 ÷ 10 = 4.030 bar.
- Find the bias
Bias = 4.030 − 4.000 = +0.030 bar. The transmitter reads high.
- Subtract it
The readings ran from 4.01 to 4.05 bar. With 0.03 bar subtracted they run from 3.98 to 4.02 bar: within 0.02 bar of the reference. Check it in the figure: press the button that subtracts the bias.
The mistake to avoid
A second trap: digits are not accuracy. A display that shows 3.903 bar from an instrument with a limit of ±0.05 bar can be wrong by 0.05 bar: its last digit carries no information, and the one before it very little.
Check yourself
A reference mass of 500 g is placed four times on a scale. The scale shows 506, 505, 507 and 506 g. What is the bias of the scale at this load?
A reference thermometer in a cold room shows 4.0 °C. Thermometer A reads 4.6, 4.6, 4.7 and 4.5 °C. Thermometer B reads 3.6, 4.5, 4.2 and 3.7 °C. Which one can be corrected so that a single reading is trusted?
An operator logs a level sensor every minute for a week. The readings are steady to the last digit. The operator concludes that the sensor needs no check against a reference. Is that right?
Where you meet it
- Two gauges on one pipeA dial gauge and a transmitter on a pump line show different pressures, and each is inside its own limit of error.
- Cold room thermometersA probe with a steady offset of half a degree can hide a real breach of the storage limit, or report one that did not happen.
- Weighing scalesA scale used for trade is checked with reference masses, because repeating a weighing cannot show its bias.
- Sensor datasheetsThe line called accuracy is a limit on the error. Repeatability, where it is given, is the precision.
- Bias is a constant offset. Spread is the scatter of repeated readings.
- Small bias is trueness, small spread is precision, and accuracy needs both.
- Averaging reduces the spread and leaves the bias unchanged.
- Only a comparison with a reference finds a bias. A known bias can be subtracted.
Discussion
No comments yet. Questions, results from your own build, a better way to explain it: all welcome.
My notes on this page
Kept on this device only. Nothing is sent anywhere.
Tip: select words in the lesson, or tap a figure or an equation, to put a note right there.
Notes on earlier wording
The lesson has changed and these words are no longer in it. Your notes are kept, with the words you chose.